Main Article Content

Abstract

A simple, selective, linear, precise and accurate RP-HPLC method was developed and validated for rapid assay of Abacavir in tablet dosage form. Isocratic elution at a flow rate of 1ml/min was employed on a symmetry C18 (250x4.6mm, 5µm in particle size) at ambient temperature. The mobile phase consisted of 0.1%orthophosphoric acid: acetonitrile: methanol 10:25:65 (V/V). The UV detection wavelength was 259nm and 20µl sample was injected. The retention time for Abacavir was 7.005 min. The method was validated as per the ICH guidelines. The method was successfully applied for routine analysis of Abacavir in tablet dosage form and in serum.

Keywords

Abacavir UV detection serum recovery precise C18 Column

Article Details

How to Cite
Ramakotaiah M, Chandrashekar K.B, & Babu Rao CH. (2011). Development and validation method for quantification of Abacavir in formula-tion and serum analysis by using RP-HPLC. International Journal of Research in Pharmaceutical Sciences, 2(1), 95-99. Retrieved from https://pharmascope.org/index.php/ijrps/article/view/811